Highly accurate and optimal |
||||||||||||||||||||||||||||||
VersionsMeasurement systems according to your needs |
||||||||||||||||||||||||||||||
FunctionsHigh precisionThe 64-KTM achieves a high precision of 0.2 K through specially developed thermal blocks with integrated cold junction compensation. The special connectors with integrated cold junction compensation suppress fluctuations in ambient temperature and guarantees long-term measurement stability.
The 64-KTM has 4 thermal blocks each with 16 universal and galvanically isolated inputs for temperature measurement. The following thermocouple types are supported: B, C, E, J, K, L, N, R, S, T, U.
Each thermal block has an integrated 4-wire RTD (class A) sensor for measuring comparable temperatures. The special mechanical construction, calibration and cold junction compensation delivers precision of 0.2 K.
The measured thermoelectric voltage at the connectors is converted and saved directly into degrees Celsius using temperature curves within the device. Measurement transducers are therefore unnecessary. The internal data memory has a capacity for 128 million data records, which equates to around a 70-day recording period. |
||||||||||||||||||||||||||||||
SoftwareJust completeThe powerful ProfiSignal software is available for data evaluation and system configuration. Extensive analysis functions within ProfiSignal enable the very easy evaluation of measurement data using different y(t), x(y) and logic diagrams. Data export as .csv files is also possible. Trend diagrams enable the user to smoothly switch, without interruption, between current online data and archived measurement data.
The ProfiSignal software enables any measurement channel to be quickly and simply set to the required thermocouple type and sampling rate. ProfiSignal's integrated DataService Configurator delivers a high degree of data security through online archiving and a user management/password system. Many customers value the 64-KTM's channels for threshold value calculation and event processing. These channels enable the computation of characteristic values during measurement runs. When required, any 64-KTM channels can be assigned an alarm function. The DataService Configurator can be used for setting threshold value monitoring conditions and event control. The ProfiSignal software also supports drivers such as LabVIEW™, OCX API and OPC-Server. |
||||||||||||||||||||||||||||||
Technical SpecificationsOverview of Informations
|
||||||||||||||||||||||||||||||