ATM/T - Analog Transmitter |
|
|
|
OVERVIEW
Features
-
Pressure measuring range: 100 mbar…1000 bar
-
Accuracy: ≤ ± 0.50 / 0.25 / 0.10 % FS
-
Temperature measuring range: -50…150 °C
-
Operating temperature: -25…85 °C
-
Process temperature: -40…150 °C
-
2 Output signals (p & T): 0…10 V DC, 0…20 / 4…20 mA
-
Materials: Stainless steel, Titanium
-
Customer benefits
-
Pressure measuring range: 100 mbar…1000 bar
-
Accuracy: ≤ ± 0.50 / 0.25 / 0.10 % FS
-
Temperature measuring range: -50…150 °C
-
Operating temperature: -25…85 °C
-
Process temperature: -40…150 °C
-
2 Output signals (p & T): 0…10 V DC, 0…20 / 4…20 mA
-
Materials: Stainless steel, Titanium
|
Back to Top |
SPECIFICATIONS
|
0.1 ... 0.5, (1) |
> 0.5 ... 2 |
> 2 ... 25 |
Overpressure |
3 bar |
3 x FS (≥ 3 bar) |
3 x FS |
Burst pressure, (5) |
≥ 30 bar |
≥ 30 bar |
3 x FS (≥ 25 bar) |
Accuracy, (6), (± % FS) |
≤ 0.5 / ≤ 0.25 |
≤ 0.5 / ≤ 0.25 / ≤ 0.1 |
≤ 0.5 / ≤ 0.25 / ≤ 0.1 |
Thermal shift, (± % FS/°C) |
|
|
|
Zero point 0 ... 70°C |
≤ 0.06 |
≤ 0.03 |
≤ 0.015 |
Zero point -25 ... 85°C |
≤ 0.08 |
≤ 0.04 |
≤ 0.02 |
Span 0 … 70°C |
≤ 0.015 |
≤ 0.015 |
≤ 0.015 |
Span -25 … 85°C |
≤ 0.02 |
≤ 0.02 |
≤ 0.02 |
Response time, (typ.) |
< 1ms / 10 ... 90% FS |
< 1ms / 10 ... 90% FS |
< 1ms / 10 ... 90% FS |
Long term stability, (7) |
< 0.5% FS / < 4 mbar |
< 0.2% FS / < 4 mbar |
< 0.1% FS / < 0.2% FS |
|
> 25 ... 600, (2), (4) |
> 600 ... 1000, (2), (3) |
Overpressure |
3 x FS (≤ 850 / ≤ 1500 bar) |
1500 bar |
Burst pressure, (5) |
> 850 / ≤ 1500 bar |
> 1500 bar |
Accuracy, (6), (± % FS) |
≤ 0.5 / ≤ 0.25 / ≤ 0.1 |
≤ 1 / ≤ 0.5 / ≤ 0.25 |
Thermal shift, (± % FS/°C) |
|
|
Zero point 0 ... 70°C |
≤ 0.015 |
≤ 0.015 |
Zero point -25 ... 85°C |
≤ 0.02 |
≤ 0.02 |
Span 0 … 70°C |
≤ 0.015 |
≤ 0.015 |
Span -25 … 85°C |
≤ 0.02 |
≤ 0.02 |
Response time, (typ.) |
< 1ms / 10 ... 90% FS |
< 1ms / 10 ... 90% FS |
Long term stability, (7) |
< 0.1% FS / < 0.2% FS |
< 0.1% FS / < 0.2% FS |
|
|
Standard, (1) |
-25 ... 100°C |
Upper end of range |
150 °C |
Lower end of range |
-40 °C |
Temperature span, (2) |
> 30 °C |
Accuracy, (3) |
|
0...70 °C, (typ./max.) |
± 0.5 / 1.0 °C |
-25...100 °C, (typ./max.) |
± 2.0 °C |
-25...85 °C, (typ./max.) |
± 1.0 / 1.5 °C |
|
|
Operating temperature |
-25 ... 85°C |
Process temperature |
-40 ... 150°C |
Storage temperature |
-25 ... 85°C |
|
4 ... 20 mA |
0 ... 20 mA |
0 ... 5 V / 0 ... 10 V |
Power supply |
15 ... 30 VDC |
15 ... 30 VDC |
15 ... 30 VDC |
Supply influence |
< 0.1% FS |
< 0.1% FS |
< 0.1% FS |
Current consumption |
|
|
|
Circuit diagram |
|
|
|
Load resistance |
(Usupply - 5V) / 0.02A |
(Usupply - 5V) / 0.02A |
RL > 10kOhm |
Load influence |
< 0.1% FS |
< 0.1% FS |
< 0.1% FS |
|
Back to Top |
CERTIFICATES
CE Declaration of Confirmity 2016-04-13
RoHS
|
|
|
|