AX/S Series Special Sensor |
|
Axial Cable Outlet (LVDT) |
A6G/1/S |
AX/0.25/S |
AX/0.5/S |
|
Axial Cable Outlet (Half Bridge) |
A6G/1/SH |
AX/0.25/SH |
AX/0.5/SH |
|
Body Diameter |
6h6 |
8h6 |
|
Measurement Range (mm) |
± 1.0 |
± 0.25 |
± 0.5 |
|
Pre-Travel (mm) |
0.15 |
0.03 |
0.15 |
|
Over Travel (mm) |
0.35 |
0.03 |
0.35 |
|
Accuracy1 |
0.5, 1.0 µm |
0.5, 0.5 µm |
0.5, 1.0 µm |
|
Repeatability (µm) |
<0.15 |
<0.1 |
<0.15 |
|
Tip Force (N) @ Centre Travel ± 20% |
0.7 |
|
Resolution |
Dependent on Electronics |
|
Temperature Coefficient (%FSO/°C) |
0.01 |
|
1Accuracy is % of reading or µm, whichever greater. |
|
Electrical Interface (LVDT Plugged) |
|
Sensitivity2 (mV/V/mm) ± 0.5% |
200 |
133 |
80 |
|
Energising Current (mA/V) ± 5.0% |
1.8 |
2.0 |
2.0 |
|
Electrical Interface (LVDT Unplugged) |
|
Sensitivity (mV/V/mm) ± 5.0% |
210 |
150 |
150 |
|
Electrical Interface (Half Bridge Plugged) |
|
Sensitivity2 (mV/V/mm) ± 0.5% |
73.5 |
49.0 |
29.4 |
|
Energising Current (mA/V) ± 5.0% |
1.0 |
1.0 |
1.0 |
|
Electrical Interface (Half Bridge Unplugged) |
|
Sensitivity (mV/V/mm) ± 5.0% |
83 |
150 |
150 |
|
2Linearity and sensitivty error combined does not exceed a total of 0.5%. |
|
Environmental |
|
Storage Temperature (°C) |
-40 to +100 |
|
Operating Temperature (°C) |
+5 to +80 |
|
IP Rating |
IP 65 |
|
|
|
Materials |
|
Case |
Stainless Steel |
|
Probe Tip |
Nylon or Tungsten Carbide |
|
Gaiter |
None |
|
Cable |
2 m PUR |