Clamping Collet |
Ø9.5 mm |
Ø10 mm |
805048 (M4 Screw)
806788 (M3 Screw)
|
806460-SX (M4 Screw) |

|

|
For use with all 8 mm clamping shaft sensors. The clamping collet distributes the clamping forces evenly around the shaft. Using the supplied grub screw, the probe can be loosened while holding the collet in place. |
Imperial Adapter Sleeves |

Adapter Sleeves can be used to increase the body diameter of 8 mm sensors to 9.512 (3/8"). Available in lengths A from 12 to 127 mm.
|
Imperial Adapter Sleeves
(Split) |

Split Adapter Sleeves can be used to increase the body diameter of 8 mm sensors to 9.512 (3/8"). Available in lengths A from 12 to 127 mm.
|
|
|
|
 |
|
|
Ultra Feather Touch Pneumatic Digital Gauging Probe (DW/P) - Specification
Products (Dia 8h6) |
Spring Push Axial Cable |
DW/10/S
|
Pneumatic Axial Cable |
DW/10/P
|
Measurement Performance |
Measurement Range (mm) |
10
|
Accuracy (% of Reading) (Note 1) |
0.06
|
Accuracy (% of Reading) (Note 1) - with in line connector |
0.15
|
Repeatability (worst case) µm (Note 2) |
0.15
|
Repeatability (typical) µm (Note 3) |
0.05
|
Resolution (µm) |
0.01
|
Pre Travel (mm) |
0.15
|
Post Travel (mm) |
0.85
|
Tip Force (N) at Middle of Range ±20% |
|
Spring Push |
0.03 to 0.06
|
Pneumatic at 0.4 bar |
0.03 to 0.06
|
Temperature Coefficient %FS/ºC |
0.01
|
Measurement Performance |
Sealing for probe |
IP50
|
Sealing for probe interface electronics |
IP43 for module and TCON
|
Storage Temperature (°C) |
-20 to +80
|
Probe Operating Temperature |
-10 o +80
|
Electronics Operating Temperature (°C) |
0 to 60
|
EMC Emissions |
EN61000-6-3
|
EMC Immunity |
EN61000-6-2
|
Probe Life |
> 10 million cycles with minimum side load applied.
|
Material |
Probe Body |
Stainless Steel
|
Probe Tip (options) |
Nylon, Ruby, Silicon Nitride, Tungsten Carbide
|
Cable |
PUR
|
Electronics Module |
ABS
|
Electronics Interface (Orbit® 3) |
Orbit® 3 Interface Options |
USB, Ethernet, RS232
|
Reading Rate |
3906 readings per second
|
Bandwidth of Electronics (Hz) user selectable |
460, 230, 115, 58, 29, 14, 7, 4
|
Note 1: Accuracy 0.1 μm or % reading whichever greater
Note 2: Repeated operation against a carbide target with side load applied to the bearing using max-min
Note 3: Repeated operation against a carbide target standard deviation from average (68%)
|
|
|
|
|
|
|
Digital Displacement Sensors | Low Tip Force | Gauge Probes
The Ultra Feather Touch probe has so light a tip force, it is a viable alternative to a non contact sensor in many applications. With various tips in ruby and nylon the UFT is already being used to check glass, rubber, semiconductor wafers and other delicate materials.
Combining multiple sensors using the Orbit Digital Measurement network is easy and data can be read using a Solartron readout or the Gauge Computer Software making system integration simple.
Attention to detail in both design and manufacture of the displacement sensor ensures precision measurement throughout the linear displacement transducer's life.
Solartron Metrology offers a large range of displacement sensors to suit multiple applications with different cable outlets (radial or axial), multiple measuring ranges, different cable types like steel braided which provides additional protection for displacement sensors that are being used in harsh environments.
|
|
|
|
|
 |
|
|
Replacement Gaiters |

Gaiters can be replaced when damaged. Only pneumatic push probes require gaiter rings.
Part Numbers:
Spring Push |
|
Part Number |
Pneumatic Push |
Part Number |
A6G/1/S |
D6P/2/S |
205014 |
|
|
|
AX/1/S |
DP/2/S |
204851 |
AX/1/P |
DP/2/P |
802691 |
AX/1.5/S |
|
204851 |
|
|
|
AX/2.5/S |
DP/5/S |
204894 |
AX/2.5/P |
DP/5/P |
802692 |
AX/5/S |
DP/10/S |
204860 |
AX/5/P |
DP/10/P |
802693 |
AX5/1/S |
DP10/2/S |
204860 |
AX5/1/S |
DP10/2/P |
802693 |
|
DP/12/S |
209562 |
|
DP/12/P |
|
AX/10/S |
DP/20/S |
205906 |
AX/10/P |
DP/20/P |
803235 |
|
|
|
|
|
|
|
LE/12/S |
209562 |
|
LE/12/P |
805098 |
|
LE/25/S |
205906 |
|
LE/25/P |
804241 |
|
Clamping Collet |
Ø9.5 mm |
Ø10 mm |
805048 (M4 Screw)
806788 (M3 Screw)
|
806460-SX (M4 Screw) |

|

|
For use with all 8 mm clamping shaft sensors. The clamping collet distributes the clamping forces evenly around the shaft. Using the supplied grub screw, the probe can be loosened while holding the collet in place. |
Probe Tips |
Various probes are available. See Probe tips |
Imperial Adapter Sleeves |

Adapter Sleeves can be used to increase the body diameter of 8 mm sensors to 9.512 (3/8"). Available in lengths A from 12 to 127 mm.
|
Imperial Adapter Sleeves
(Split) |

Split Adapter Sleeves can be used to increase the body diameter of 8 mm sensors to 9.512 (3/8"). Available in lengths A from 12 to 127 mm.
|
|
|
|
 |
|
|
Feather Touch Pneumatic Digital Gauging Probe (DT/P) - Specification
Feather Touch |
Axial Cable Outlet |
DT/2/P |
DT/5/P |
DT/10/P |
DT/20/P |
DT10/2S |
Radial Cable Outlet |
DTR/2/P |
DTR/5/P |
DTR/10/P |
DTR/20/P |
DTR10/2S |
|
Measurement Performance |
Measurement Range (mm) |
2 |
5 |
10 |
20 |
2 |
Accuracy (% of Reading) (Note 1) |
0.05 |
0.05 |
0.06 |
0.07 |
0.05 |
Accuracy (% of Reading) (Note 1) - with In Line Connector |
0.2 |
0.15 |
0.15 |
0.15 |
0.2 |
Repeatability (worst case) µm (Note 2) |
0.15 |
0.15 |
0.15 |
0.15 |
0.15 |
Repeatability (typical) µm (Note 3) |
0.05 |
0.05 |
0.07 |
0.07 |
0.05 |
Resolution (µm) |
0.01 |
0.05 |
0.05 |
0.1 |
0.01 |
Pre Travel (mm) |
0.15 |
0.15 |
0.15 |
0.15 |
0.15 |
Post Travel (mm) |
0.85 |
0.85 |
0.85 |
0.85 |
8.85 |
Tip Force (N) at Middle of Range ±20% |
|
|
|
|
|
Pneumatic at 0.3 Bar |
0.18 |
0.18 |
0.18 |
0.18 |
0.18 |
Pneumatic at 1 Bar |
1.1 |
1.1 |
1.1 |
1.1 |
1.1 |
Temperature Coefficient %FS/°C |
0.01 |
0.01 |
0.01 |
0.01 |
0.01 |
|
Environmental |
Sealing for Probe |
IP65 with gaiter or IP50 without gaiter |
Sealing for Probe Interface Electronics |
IP43 for module and TCON |
Storage Temperature (°C) |
-20 to +80 |
Probe Operating Temperature with Gaiter (°C) |
-5 to +80 |
Probe Operating Temperature without Gaiter (°C) |
-10 to +80 |
Electronics Operating Temperature (°C) |
0 to +60 |
EMC Emissions |
EN61000-6-3 |
EMC Immunity |
EN61000-6-2 |
Probe Life |
100 million cycles (no side load), > 10 million cycles in most applications |
|
Material |
Probe Body |
Stainless Steel |
Probe Tip (options) |
Nylon, Ruby, Silicon Nitride, Tungsten Carbide |
Gaiter (Note 6) |
Fluoroelastomer or Silicon |
Cable |
PUR |
Electronics Module |
ABS |
|
Electronics Interface (Orbit®3) |
Orbit®3 Interface Options |
USB, Ethernet, RS232 |
Reading Rate |
3906 readings per second |
Bandwidth of Electronics (Hz) user selectable |
460, 230, 115, 58, 29, 14, 7, 4 |
Power |
5±0.25 VDC @ 0.06A typical |
|
1Accuracy 1 µm or % reading, whichever greater. |
2Repeated operation against a carbide target with side load applied to the bearing using max-min. |
3Repeated operation against a carbide target standard deviation from average (68%) |
6Different gaiter materials available for specific applications - Fluoroelastomer standard option |
|
|
|
|