| Analog inputs |
4, 8, 12 or 16 synchronous sampling |
| Sampling rate, adjustable per channel |
20 Hz .. 50 kHz |
| Usable signal range |
DC .. 20 kHz |
| Voltage range |
± 25 V |
| Current measuring range |
0 .. 20 mA / 4 .. 20 mA / free |
| Signal conditioning, software switchable |
None, AC coupling, IEPE |
| Resolution/type ADC / input impedance |
24 bit Sigma-Delta / 4 GΩ |
| Dielectric withstand voltage |
± 100 VDC |
| Galvanic isolation channel to channel / channel to system / PE |
± 400 VDC |
| Measurement accuracy |
0.5 mV + 0.008 % from measurement value |
| Digital inputs |
4 |
| Input signal |
low: 0 .. 1 V / high: 5 .. 100 VDC@3.5 mA |
| Galvanic isolation channel to channel / channel to system / PE |
± 400 VDC |
| Max. input frequency / min. pulse width |
1 MHz / 500 ns |
| Analog outputs |
4 |
| Resolution |
16 bit |
| Output ranges |
0 .. 10 V / ± 10 V / 0 .. 20 mA / 4 ... 20 mA / ± 20 mA |
| Galvanic isolation channel to other channel types / Duct to system / PE |
± 400 VDC |
| Min. load / max. burden |
650 Ω |
| Digital outputs |
8 |
| Max. switching voltage / current |
50 V / 2.5 A for 2 seconds, 250 mA continuous current |
| Galvanic isolation channel to other channel types / Duct to system / PE |
± 400 VDC |
| PWM basic frequency / duty cycle |
5 Hz .. 10 KHz / 1:1000 |
| Data storage |
| Internal / external |
32 GB (approx. 2,300 million measured values) / USB, NFS, CIFS, (S)FTP |
| Signal conditioning functions |
| High-pass / low-pass / bandpass filter |
| Cut-off frequency / filter order / filter characteristic |
5 .. 20 kHz / 4, 6, 8, 10 / Butterworth; Chebyshev, Bessel, Critical Damping |
| Analysis |
| Integrator / Differentiator |
Single or double integrator / differentiator |
| FFT |
| Line number / window function / averaging |
100 .. 12.800 lines (zero padding up to 32k points) / von-Hann, Hamming, Flat-Top, Tukey / 2 .. 32 times |
| FFT types |
Narrow band / wide band, envelope / demodulation, amplitude / phase spectral |
| Characteristic values from time signal |
| |
Maximum / minimum value, peak-to-peak value, arithm. average, RMS, max. vect. sum, air-gap, crest factor |
| Characteristic values from frequency spectrum |
| |
Amplitude, frequency and phase of main oscillation and any harmonics, sum value, quadratic mean (in any frequency band), value remainder |
| |
Amlitudes for various damage frequencies of rolling bearing, gear stage, planetary stage |
| Characteristic values on digital input |
| Frequency |
0.2 Hz .. 1 MHz |
| Counter |
64 bit, UP-Down counter, quadratur decoder |
| Interfaces |
| Physical equipment COM 1 / COM 2 |
RS485, 9-pole sub-D socket, IEC 61158-2 |
| Physical equipment COM 3 |
RS232, 9-pole sub-D-plugs |
| LAN |
1 x 1000BASE-T |
| USB |
Device 2.0 / Host 2.0 / low / high / full |
| PROFIBUS |
2 x PROFIBUS DPV1 / Slave max. 12 Mbit, also redundant according to PNO 2.212 V1.2 |
| CAN / RS 232/485 |
2 x CAN 2.0 / Modbus RTU Master & Slave, SCPI, ASCII |
| TCP/IP |
Modbus TCP Server & Client, OPC UA Server & Client, OPC UA HA Server (optional), SMTP, NTP, PTP |
| PROFINET (TCP/IP) |
PROFINET RT Device, IO/MRP/LLDP, Conf Class B, RT Class 1, Netload Class III, I&M 0,1,2,3,5, 2xRJ45 100MBit/s full duplex |
| General technical information |
| Dimensions / weight |
W 217 mm x H 88 mm x D 128 mm / 750 g |
| Fixing |
Mounting rail DIN EN 60715 or screw mounting |
| Connecting terminals |
Plug-in screw terminals, 96 terminals in 2 rows, max. 1.5 mm² |
| Temperature range |
-20 .. 50 °C* |
| Humidity |
max. 90 % relative humidity, non-condensing |
| Supply voltage |
12 .. 24 VDC / ± 10% |
| Power input |
Max. 20 watts |
|
| * Sufficient air circulation in the control cabinet must be ensured. |
| |
| |
|