DTM.OCS.S
 

DOWNLOADS

 

 

OVERVIEW

 

Features

  • Pressure measuring range: 200 mbar…100 bar
  • Accuracy: ≤ ± 0.15 / 0.05 / 0.03 % FS
  • Operating temperature: -40…85 °C
  • Process temperature: -40…85 °C
  • Interface: RS485
  • Materials: Stainless steel, Titanium
  •  

Customer benefits

  • Pressure measuring range: 200 mbar…100 bar
  • Accuracy: ≤ ± 0.15 / 0.05 / 0.03 % FS
  • Operating temperature: -40…85 °C
  • Process temperature: -40…85 °C
  • Interface: RS485
  • Materials: Stainless steel, Titanium
Back to Top

SPECIFICATIONS

Pressure measuring range (bar)

  0.2 ... 0.5 > 0.5 ... 2 > 2 ... 100
Overpressure 3 bar 3 x FS (≥ 3 bar) 3 x FS
Burst pressure, (1) > 200 bar > 200 bar > 200 bar
Accuracy, (2), (± % FS) ≤ 0.15 ≤ 0.05 ≤ 0.03
Total Error, (3) (4) (± % FS)      
-40...85°C (typ. / max.) ≤ 0.8 / 1.5 ≤ 0.3 / 0.5 ≤ 0.1 / 0.2
-25...85°C (typ. / max.) ≤ 0.5 / 1.0 ≤ 0.2 / 0.4 ≤ 0.1 / 0.2
0...70°C (typ./max.) ≤ 0.3 / 0.8 ≤ 0.1 / 0.2 ≤ 0.1 / 0.2
Long term stability, (5) < 0.5% FS / < 4 mbar < 0.2% FS / < 4 mbar < 0.1% FS / < 0.2% FS

Temperature measuring range, (1) (°C)

  0 ... 70 -25 ... 85 -40 ... 85°C
Accuracy, (2) ≤ ± 0.5 °C ≤ ± 1.5 °C ≤ ± 1.5 °C
Response time, (3), (4)      
T 0.50 9 s 9 s 9 s
T 0.90 27 s 27 s 27 s
T 0.63 15 s 15 s 15 s

Temperature range (1)

   
Operating temperature -40...85 °C
Process temperature -40...85 °C
Storage temperature -40...85 °C

Electrical specifications

   
Power supply (1) 9...30 VDC
Current consumption (typ.) 3mA @ 24VDC
Resolution  
Temperature 0.1°C
Pressure 20 Bit
Output  
Baudrate 9600 bps
Protocol Modbus RTU, 8n2
Interface RS485
Cable length (2) max. 1000m

Physical specifications

 
 

 