LDM-Series - Laser Micrometer Display Units

 

SPECIFICATIONS

2000/400 Scan type
  LDM-110 LDM-110EX LDM-1000 LDM-1000EX
 Picture
 No. of Axes For Single Max. 2 Max. 5 Max. 3
 Scanning cycle 400 times/sec 400 times/sec 400 times/sec 400 times/sec
 No. of Averaging time 1-2048
(16 steps)
1-2048
(8 steps)
1-2048(22 steps) at LDM-1000LS
1-1024(19 steps) at LDM-1000XY
 Measuring mode 16 kinds 8 kinds
(X,Y axis each)
16 kinds + Optional edges measurement
per each axis
 Processing function NG Average, Difference, Large dia. measurement, etc.
 Cable cover(Wall) thickness processing NG NG OK(by LDM-1000XY)
 Display device 7 segment LED Graphic LCD 7 segment LED
 No. of Display digit 4-5 digits 4-5 digits 3-6 digits 3-6 digits
 Processing contents monitor NG NG OK OK
 Center position monitor LED(8pcs) LED(8pcs)
(X,Y axis each)
Graphic bargraph
per each axis
LED(8pcs) per each axis
No. of channel of Recorder output 1 1 1 channel per board(LDM-1000LS/XY)
 Recorder output performance Max.+/-10V, Resolution 12bits(=approx.10mV/1bit), 
Output impedance 100ohm(approx.),
Deviation/Absolute value switch over available
 Recorder output gain 1μm-1m/V(decade steps) 0.1μm~1m/V(decade steps)
 Digital dia. controller connection OK OK NG NG
 Center value setting 5 digits 5 digits 6 digits 6 digits
 High/Lo limit tolerance setting 4 digits 4 digits 6 digits 5 digits
 High/Low limit contact output 1a contact(*2) 1a contact(*2) per board(LDM-1000LS/XY)
Contact rating: AC250V 2A(Resistance load)
 Error contact output 1b contact(Close when power OFF) 1b contact(Close when power OFF)(Combined use by each board)
Contact rating: AC250V 2A(Resistance load)
 Laser remote terminal OK OK (X,Y common) OK(each LDM-1000LS board)
 Hold terminal OK OK (X,Y,P common) OK(each LDM-1000LS/XY board)
 Edge monitoring function OK OK OK OK
 Self-diagnosis mode OK OK OK OK
 Multi-point calibration function OK NG NG NG
 Trigger function (Max. meas.,  Deviation meas., etc.) NG NG OK OK
 Offset function NG NG OK OK
 Zero set function OK NG OK OK
 mm/inch changeover function NG NG OK OK
 Programing function NG NG OK(30 kinds) OK(9 kinds)
 Graph display NG NG OK NG
 Serial port OK OK OK OK
 BCD output Option OK(by LDM-1000BCD)
 RS-485 Function Option NG
 Cable length Standard 5 meters, Max. 30 meters (Extensible to 120m.max. as option)
 Power supply AC100-240V Free power supply 50/60Hz
 Power consumption less than 11VA less than 14VA less than 15VA(only display unit)
less than 8VA(each board)
 Ambient condition Temperature:0-45 degree centigrade, Humidity:35~85%RH(no condensation)
 Outside dimension 144(W) * 72(H) * 191(D)[mm] 216(W) * 144(H) * 225(D)[mm]
 Weight Approx. 1.5kg Approx. 1.6kg Approx. 4kg (only display unit)
Approx. 280g (each board)
 
2000/400 Scan type
  LDM-210 LDM-210C LDM-210EXC
 Picture
 No. of Axes For Single Max. 2
 Scanning cycle 2000 times/sec or 400 times/sec
 No. of Averaging time 1-2048 (12 steps)
 Measuring mode Optional edges measurement per each axis
 Processing function NG Average, Difference, Large dia. measurement, etc.
 Cable cover (Wall) thickness processing NG
 Display device 7 segment LED
 No. of Display digit 4-5 digits
 Processing contents monitor NG
 Center position monitor LED(8pcs) LED(15pcs) LED(8pcs) (X,Y axis each)

 No. of channel of recorder output

1
 Recorder output performance Max.+/-10V, Resolution 12bits(=approx.10mV/1bit), 
Output impedance 100ohm(approx.),
Deviation/Absolute value switch over available
 Recorder output gain 1μm-1m/V(decade steps)
 Digital dia. controller connection OK NG
(Built-in Dia.Controller)
 Center value setting 5 digits
 High/Lo limit tolerance setting 5 digits

 High/Low limit contact output

1a contact(*2)
Contact rating: AC250V 2A(Resistance load)
 Error contact output 1b contact(Close when power OFF)
Contact rating: AC250V 2A(Resistance load)
 Laser remote terminal OK OK(X,Y common)
 Hold terminal OK OK(X,Y common)
 Edge monitoring function OK
 Self-diagnosis mode OK
 Multi-point calibration function OK

 Trigger function (Max. meas., Deviation meas., etc.)

NG

 Offset function

OK(Ver2.15 or more) NG
 Zero set function OK NG
 mm/inch changeover function NG
 Programing function NG
 Graph display NG
 Serial port OK
 BCD output Option
 RS-485 Function Option
 Cable length Standard 5 meters, Max. 30 meters (Extensible to 120m.max. as option)
 Power supply AC100-240V Free power supply 50/60Hz
 Power consumption less than 15VA less than 20VA
 Ambient condition Temperature:0-45 degree centigrade, Humidity:35~85%RH(no condensation)
 Outside dimension 144(W) * 72(H) * 191(D)[mm] 144(W) * 72(H) * 241(D)[mm]
 Weight Approx. 1.5kg Approx. 1.8kg

ADD-ONS

 

Signal Extender(LDM-201, LDM202) - extends cables when the distance between a scanning unit and a display unit exceeds 30    metres.

Signal Distributor(LDM-205, LDM-206) - divides output signals from a scanning unit.

BCD Distributor(LDM-B2)

 
 

 

Product & Technology News

LDM Series

Download Catalogue