Standard |
Axial Cable Outlet |
DP/2/P |
DP/5/P |
DP/10/P |
DP/20/P |
DP10/2/P |
Radial Cable Outlet |
DPR/2/P |
DPR/5/P |
DPR/10/P |
DPR/20/P |
DPR10/2/P |
|
Measurement Performance |
Measurement Range (mm) |
2 |
5 |
10 |
20 |
2 |
Accuracy (% of Reading) (Note 1) |
0.05 |
0.05 |
0.06 |
0.07 |
0.05 |
Accuracy (% of Reading) (Note 1) - with In Line Connector |
0.2 |
0.15 |
0.15 |
0.15 |
0.2 |
Repeatability (worst case) µm (Note 2) |
0.15 |
0.15 |
0.15 |
0.15 |
0.15 |
Repeatability (typical) µm (Note 3) |
0.05 |
0.05 |
0.07 |
0.07 |
0.05 |
Resolution (µm) |
0.01 |
0.05 |
0.05 |
0.1 |
0.01 |
Pre Travel (mm) |
0.15 |
0.15 |
0.15 |
0.15 |
0.15 |
Post Travel (mm) |
0.85 |
0.85 |
0.85 |
0.85 |
8.85 |
Tip Force (N) at Middle of Range ±20% |
|
|
|
|
|
Pneumatic at 0.4 Bar |
0.7 |
0.7 |
0.7 |
0.7 |
0.7 |
Pneumatic at 1 Bar |
2.6 |
2.6 |
2.6 |
2.6 |
2.6 |
Temperature Coefficient %FS/°C |
0.01 |
0.01 |
0.01 |
0.01 |
0.01 |
|
Environmental |
Sealing for Probe |
IP65 with gaiter or IP50 without gaiter |
Sealing for Probe Interface Electronics |
IP43 for module and TCON |
Storage Temperature (°C) |
-20 to +80 |
Probe Operating Temperature with Gaiter (°C) |
-5 to +80 |
Probe Operating Temperature without Gaiter (°C) |
-10 to +80 |
Electronics Operating Temperature (°C) |
0 to +60 |
EMC Emissions |
EN61000-6-3 |
EMC Immunity |
EN61000-6-2 |
Probe Life |
100 million cycles (no side load), > 10 million cycles in most applications |
|
Material |
Probe Body |
Stainless Steel |
Probe Tip (options) |
Nylon, Ruby, Silicon Nitride, Tungsten Carbide |
Gaiter (Note 6) |
Fluoroelastomer or Silicon |
Cable |
PUR |
Electronics Module |
ABS |
|
Electronics Interface (Orbit®3) |
Orbit®3 Interface Options |
USB, Ethernet, RS232 |
Reading Rate |
3906 readings per second |
Bandwidth of Electronics (Hz) user selectable |
460, 230, 115, 58, 29, 14, 7, 4 |
Power |
5±0.25 VDC @ 0.06A typical |
|
1Accuracy 1 µm or % reading, whichever greater. |
2Repeated operation against a carbide target with side load applied to the bearing using max-min. |
3Repeated operation against a carbide target standard deviation from average (68%) |
6Different gaiter materials available for specific applications - Fluoroelastomer standard option |