|
|
|
DL.OCS/N/RS485 - Level Logger |
|
DOWNLOADS
|
|
OVERVIEW
Features
-
Pressure measuring range: 0 … 250 mH2O
-
Total Error: ≤ ± 0.05 % FS (-5 ... 50°C)
-
Temperature measuring range: -5 … 80°C
-
Operating temperature: -5 … 80°C
-
Process temperature: -5 … 80°C
-
Interface: RS485
-
Materials: Stainless steel, titanium
-
Battery operation, Measurement interval 1s … 24h, 1.5 Mio. Measurement values
-
Customer benefits
-
Pressure measuring range: 0 … 250 mH2O
-
Total Error: ≤ ± 0.05 % FS (-5 ... 50°C)
-
Temperature measuring range: -5 … 80°C
-
Operating temperature: -5 … 80°C
-
Process temperature: -5 … 80°C
-
Interface: RS485
-
Materials: Stainless steel, titanium
-
Battery operation, Measurement interval 1s … 24h, 1.5 Mio. Measurement values
|
Back to Top |
SPECIFICATIONS
|
2 ... 5 |
> 5 ... 20 |
> 20 ... 250 |
Overpressure |
≥ 3 bar |
≥ 3 x FS (≥ 3 bar) |
≥ 3 x FS |
Burst pressure, (1) |
> 200 bar |
> 200 bar |
> 200 bar |
Accuracy, (± % FS) |
≤ 0.15 |
≤ 0.05 |
≤ 0.03 |
Total Error, (2), (3)
(± % FS) |
|
|
|
-5 ... 80°C, (typ. / max.) |
≤ ± 0.50 / 1.00 |
≤ ± 0.10 / 0.20 ≤ ± 1.0 / 2.0 cmH2O |
≤ ± 0.10 / 0.20 |
-5 ... 50°C, (typ. / max.) |
≤ ± 0.20 / 0.40 |
≤ ± 0.10 / 0.20 ≤ ± 0.5 / 1.0 cmH2O |
≤ ± 0.05 / 0.10 |
Long term stability, (4) |
< 0.5 % FS / < 4 mbar |
< 0.2 % FS / < 4 mbar |
< 0.1 % FS / < 0.2 % FS |
|
-5 ... 50 |
-5 ... 80 |
Accuracy (2) |
≤ ± 0.5 °C |
≤ ± 1.5 °C |
Response time, (3), (4) |
|
|
T 0.50 |
9 s |
9 s |
T 0.90 |
27 s |
27 s |
T 0.63 |
15 s |
15 s |
|
|
Operating temperature, (1) |
-5 ... 50°C / -5 ... 80°C |
Process temperature, (1) |
-5 ... 50°C / -5 ... 80°C |
Storage temperature |
-40 ... 85°C |
|
|
Data format |
Data are stored in ASCII format |
Data memory |
Up to 1.5 Mio measurement values per channel, data remains in memory even without battery, each measurement value is correlated with time and date |
Data transfer |
Read out data per measurement series, Read out all stored data, Read out data for a defined time-period |
Real-time clock |
Quartz-precision clock with date, Start-time of datalogging configurable |
Identification |
Each datalogger has a unique serial number, as well as a user-definable description |
Battery indicator |
Battery level indicator, calculated by considering the ambient temperature and the battery properties |
Configuration |
Sample rate, threshold values, Identification (f.e. measuring series), Taring of measurement value, Density of the measuring medium, Measurement units |
|
|
PC / Notebook |
Min. 1.6 GHz Dual Core x86, Memory: Min. 10 GB, RAM: Min. 2 GB |
Tablet PC |
Please contact STS |
Operating System (1) |
Windows XP SP3, 7, 8, 8.1, 8 Pro (32-/ 64-Bit), 10 |
|
Pressure |
Temperature |
PUR |
≤ 25 bar |
-5 … 50°C |
FEP |
≤ 25 bar |
-5 … 80°C |
PE |
≤ 25 bar |
-5 … 80°C |
|
Back to Top |
ACCESSORIES
|
|
|
114209 |
DL.OCS PC Application Software |
DL.OCS FW |
|
Back to Top |
CERTIFICATES
CE Declaration of Confirmity 2016-04-13
RoHS
|
|
|
|
|