Voltage and current measurement range |
± 156 mV .. ± 10 v etc. |
Sensors |
Thermocouples,RTD |
- |
Potential isolation to system |
750 VDC |
Potential isolation channels |
110 V |
650 V |
100 V |
Resolution |
24 Bit |
14 Bit |
Measurement accuracy |
V / mA: 0,1 %
RTD100: 0.1 K
RDT000: 0.05 K
Thermoelemente: 0.1 % EW
|
0.1 %
-
-
-
|
Resolution
|
16 Bit |
- |
12 Bit |
Potential isolation |
750 V |
- |
100 V |
Output |
0/4 .. 20 mA |
- |
0 .. 10 V |
Max. load / min. load |
650 Ω |
- |
2.5 kΩ |
Potential isolation |
2.5 kV |
Input measurement range |
low: 0 .. 1.5 VDC@0 .. 1.5 mA / high: 3.5 .. 90 VDC@2 mA |
Potential isolation / input measurement range |
2.5 kV / low: 0 .. 1.5 VDC@0 .. 1.5 mA / high: 3.5 .. 90 VDC@2 mA |
Measurement frequency |
bis 30 kHz bei TTL-Pegel |
Potential isolation |
2.5 kV |
Switching voltage |
max. 50 VDC@2.5 A |
Standard size / measurment values |
Storage partitionable as required, standard: 3.5 GB / max. 250 million measurement values |
Max. size / measurement values |
15.5 GB / up to 1 billion measurement values (7 billion measurement values for LogMessage 4000) |
Physical equipment COM 1 / COM 2 |
RS485, 9-pole sub-D-plug, DIN EN ISO 19245-1 |
Physical equipment COM 3 / COM 4 |
RS232, 9-pole sub-D plugs |
Protocols COM 1 .. COM 4 |
Modbus RTU master / slave, customer-specific protocols |
Ethernet |
RJ45 (8-pole STP-plug), 100 Base-TX; Protocols: TCP/IP, HTTP, SMTP, NTP, Modbus TCP Client / server |
USB |
USB 1.1. for configuration and data memory read out |
CAN |
9-pole sub-D plug, Protocols: CAN, RAW; baud rates: 50 k .. 1 MBaud |
Mechanical type |
3-pole Phoenix connector; internal bus to connect additional modules |
Baud rate / length |
1 MBaud (adjustable) / up to 10 m (1 MBaud) |
Dimension / weight |
200 x 73 x 118 mm / 1 kg |
Fixing |
Rail mounting DIN EN 60715 or screw fixing |
Signal connections |
Detachable screw terminals, 33 terminals in 2 rows, lead protection, connecting cabling max. 2.5 mm² |
Temperature range |
-20 .. 60 °C |
Power supply |
12 .. 36 VDC / 12 .. 28 VAC eff. / ± 10 %, for LM 4000 min. 18 VAC/DC |
Power input |
< 10 Watt
|