Orbit® LTH - Specification

 

Performance Specification LTH laser

 

LTHD/
25/2/B

LTHD/
50/10/B
LTHD/
50/20/B
LTHD/
120/20/B
LTHD/
120/40/B
LTHD/
200/100/B
LTHD/
300/200/B
Range (mm) 2 10 20 20 40 100 200
Offset (mm) Note 1 25 50 50 120 120 200 300
Spot Size (Diameter µm) 30 36 36 100 100 100 130
Laser Angle (degrees) 45 30 30 20 20 12 8
Linearity (±% FSO) (Note 2)  
Best (±% FSO) 0.01 0.02 0.025 0.025 0.03 0.03 0.03
Typical (±% FSO) 0.02 0.04 0.045 0.06 0.05 0.04 0.04
Best (+µm) 0.2 2 5 5 12 30 60
Typical (+µm) 0.4 4 9 12 20 40 80
Repeatability (μm) (Note 3)  
Best 0.1 0.2 0.4 0.5 1.0 3.0 7.0
Typical 0.2 0.4 0.8 1.0 2.0 6.0 15.0
Resolution (μm)  
(Note 4) 0.0076 0.0381 0.0763 0.0763 0.1526 0.3815 0.7629
(Note 5) 0.02 0.05 0.23 0.23 0.80 2.00 4.00
Laser  
Modes (Note 7) Diffuse or Specular Diffuse only
Weight Head only (g) 203 460
Power mW / Class (IEC 60825) <5 / 3R
Wavelength nm 670
Performance  
Max Sampling Frequency (kHz) 40.00
Orbit® Data Rate (Readings/sec) 3906
Sampling Cycles 256/512 μS or 1/2/4/8/16/32/64 ms (Selectable)
Working Bandwidth Hz (Note 6) 1300, 650, 325, 163, 81, 40, 20, 10 , 5
Operating Temperature °C 0 to +40
Temperature Coefficient ±0.05% of FSO/°C
Humidity 10 to 95% non condensing
Emissions/Immunity EN61000-6-3 / EN61000-6-2
Power Orbit® 3 5±0.25 VDC @ 0.09A and 24±2.5 VDC @ 0.06A typical
Power Orbit ACS 18-24 VDC @ 0.13A typical
Interfaces  
Orbit® 3 Interfaces to the Orbit®3 network via the Orbit Support Pack for Windows (for Microsoft
.Net Framework), USB, Ethernet and Serial Interface modules available
Orbit® ACS Integral Screen and Keypad, MODBUS or Serial Interfaces

 

Note 1: Distance from the laser face to the middle point of the measuring range (mm)


Note 2: Measured on white photographic paper with the laser sample rate set to 4kHz and averaging 4 ms.

Note 3: Measured on white photographic paper with the laser sample rate set to 4 kHz and averaging 16 ms, the laser beam is
              blocked between each measurement. 

Note 4: Resolution 1 LSB of the Digital System

Note 5: Standard Deviation of 25 Measurements with the laser pointing at a fixed white photographic paper target with the laser
              sample rate set to 4 kHz and averaging 16mS

Note 6: Real measurement bandwidth based on ability to reconstruct sine wave at filter frequency 

Note 7: Specular Mode is recommended for high reflective (shiny) surfaces. ND filter required specify when ordering. 

 
 

 