1For other input voltages, please contact Solartron.
2For other output frequencies, please contact Solartron.
Signal Input (Sensor Sensitivity Range)
Input Range
Up too 2.5 Vrms
Input Load Resistance (kΩ)
100
Signal Output
Voltage Output (VDC)
Up to ± 10
Maximum Current Output (mA)
11
Output Ripple
< 40 mVrms
Output Offset
100%
Temperature Coefficient Gain
< 0.03% FRO/°C
Temperature Coefficient Offset
< 0.025% FRO/°C
Warm Up Time
15 Minutes (recommended)
Linearity
< 0.1% FRO
Bandwidth-3db (Hz)
250 typical
Environmental
Operating Temperature Range (°C)
0 to +70
Storage Temperature Range (°C)
-20 to +85
IP Rating
IP 40
IP 65
Mechanical and Connections
Connections
Solder Pad or Factory Fit
Factory Fit only
Enclosure Dimensions without Connectors (mm)
98.5 x 30.5 x 13.0
73.0 x 20.6
Weight (g)
30
75
Material
ABS
400 Series Stainless Steel
Cable Lengths
All specification limits assume a nominal 3 m cable length between transducer and BICM. The BICM can be mounted up to 10 m from the sensor, but this may result in reduced performance. Not all transducers can cope with long cable lengths.
(Cable from the BICM to the processing or display unit should be limited to 100 m).